今天: 搜索:
专题
频道
  您现在的位置是:首 页 > 技术资料 > 生物安全柜

二级生物安全柜biosafety cabinet工作原理示意图
【来源:无 】 【作者: 不详】 【已经浏览3231次】
30%外排,70%内循环

Class II Type A1 BSC Characteristics
二级A1型生物安全柜特性
Exhaust approximately 30% of total air handled and recirculate 70% in cabinet
30%外排,70%内循环
Minimum inflow air velocity is 75 fpm (0.38 m/s) through the front opening
May exhaust HEPA filtered air back into the room or through a canopy (to the outdoors)
Audiovisual alarm required when connected to a building exhaust system
May have contaminated ducts and plenums under positive pressure to the room 
Type A1 cabinets are not suitable for work with volatile toxic chemical and volatile radionuclides.
 

 

 

中华人民共和国信息产业部备案号: 京ICP备05027443号-1
郑重声明:凡转载或引用本站资料须经本站许可 版权所有 :生物安全网
联系地址:北京·生物安全网 联系信箱:biosafety#qq.com (将#替换为@) 联系QQ:122556812  
Copyright © 2005 www.biosafety.com.cn设计维护:时代中广传媒